Sign in →

Test Code DTComp Drug Test Urine Comprehensive w/o Preliminary

Useful For

Detection of a wide variety of drugs in urine.

Patient Preparation

No special patient preparation required

Collection Method

Random urine

Minimum Volume

5 mL

Container

Sterile Cup

Preferred Specimen

30 mL freshly voided urine

Acceptable Specimens

5 mL freshly voided urine

Reference or Target Ranges

Cutoff Levels Comprehensive Drug Screen Panel
* EIA comprehensive screen performed on Olympus AU680
Drug Screening Cutoff
6-Acetylmorphine 10 ng/mL
Acetaminophen 25 ng/mL
Barbiturates 300 ng/mL
Benzodiazepines 300 ng/mL
Buprenorphine 5 ng/mL
Cannabinoids (THC) 50 ng/mL
Cocaine 150 ng/mL
Ethanol 20 mg/dL
Fentanyl 10 ng/mL
Opiates 300 ng/mL
Oxycodone 100 ng/mL
Phencyclidine (PCP) 25 ng/mL
Salicylate 5 ng/mL
Drug Test Urine Comprehensive Limit of Quantitation (LOQ)
* Drugs included in comprehensive drug test performed by GCMS analysis
Drug LOQ
Amphetamine 125 ng/mL
Alprazolam 20 ng/mL
Butalbital 100 ng/mL
Buprenorphine 50 ng/mL
Carisoprodol 100 ng/mL
Cocaine 20 ng/mL
Codeine 60 ng/mL
Clonazepam 20 ng/mL
Fentanyl 10 ng/mL
Hydrocodone 60 ng/mL
Hydromorphone 60 ng/mL
Methamphetamine 125 ng/mL
Meperidine 100 ng/mL
Methadone/EDDP 100 ng/mL
Morphine 60 ng/mL
Nordiazepam 20 ng/mL
Oxycodone 60 ng/mL
Oxymorphone 60 ng/mL
Oxazepam 20 ng/mL
Phenobarbital 100 ng/mL
Phencyclidine 20 ng/mL
Temazepam 20 ng/mL
Tramadol 100 ng/mL

The following drugs will have cutoff concentrations that vary depending upon which class members and/or metabolites are present. Some drugs are only detectable at toxic concentrations as the detection limits vary depending on the drug being identified:
6-Acetylmorphine, Acetaminophen, Amitriptyline, Aripiprazole, Bupropion, Caffeine, Carbamazepine, Chlorpheniramine, Gabapentin, Hydroxyzine, Ibuprofen, Imipramine/Desipramine, Ketamine, Lamotrigine, Levetiracetam, Lidocaine, Meprobamate/Carisoprodol, Methylphenidate, Metoprolol, Mirtazapine, Naproxen, Nicotine/Cotinine, Nortriptyline, Oxcarbazepine, Paroxetine, Phentermine, Phenytoin, Quetiapine, Quinine and/or Quinidine, Salicylate, Sertraline, Topiramate, Valproic Acid, Venlafaxine

Reportable Units

ng/mL

Critical Value

Not defined for this assay

Reasons for Rejection

 


<!-- Rejection-->

 

Hemolysis Threshold N/A
Icterus Threshold N/A
Lipemia Threshold N/A
Other QNS, contaminated, improperly labeled, improper specimen type

Specimen Stability

 


<!-- Stability -->

 

Ambient 7 days
Refrigerated (4°C to 8°C) 14 days
Frozen(-40°C to 0°C) 365 days

Performance Information

 


<!-- Performance information -->

 

Days and Time Performed Monday – Friday 730 AM – 430 PM
Expected Turn Around Time 24 – 48 hours
Stat Availability No
Performing Bench Toxicology
Methodology/Method Description Screen: Olympus AU680 EIA
Confirmation: Liquid/Liquid extraction, Solid phase extraction, Gas chromatography mass spectrometry

Additional Information/Important Notes

Patients that fall below the cutoff will be reported as negative.

CPT Codes

80307, G0483

Panel Components

See listing under reference ranges