Sign in →

Test Code CBC Man CBC with Manual Differential

Important Note

This test has been discontinued. Please refer to CBC Auto.

Aliases

HP with Manual Differential

Useful For

A screening tool to confirm a hematologic disorder, to establish or rule out a diagnosis, to detect an unsuspected hematologic disorder, or to monitor effects of radiation or chemotherapy. These counts can be used as clinical guides in the diagnosis or monitoring of many diseases.

Collection Method

Venous draw, Capillary Draw

Minimum Volume

Adult: 1 mL, Pediatric 300 uL

Container

Purple (EDTA)

Preferred Specimen

Whole Blood

Acceptable Specimens

EDTA Whole Blood, BD EDTA MAP tube

Reference or Target Ranges

Click here for reference ranges

Critical Value

Parameter Low value High value
ANC (any age) <0.5 X 10 9/L N/A
WBC (0-18 years old) <2 X 109/L >50 X 109/L
WBC (18 and greater years old) <1 X 109/L >80 X 109/L
Hemoglobin (8 weeks and up) <6 g/dL >21 g/dL
Hematocrit (any age) <21% >65%
Hematocrit (0-24 hours) <42% >65%
Platelets (any age) <20 X 109/L >1,000 X 109/L

Reasons for Rejection

Hemolysis Threshold N/A
Icterus Threshold N/A
Lipemia Threshold N/A
Other Clotted; QNS, Contaminated; improperly labeled

Specimen Stability

Ambient Up to 24 hrs. at room temp; Slide must be made and stained within 8 hrs.
Refrigerated (4°C to 8°C) 24 Hours
Frozen(-70°C to 0°C) N/A

Performance Information

Days and Time Performed 24/7
Expected Turn Around Time Stat: CBC (w/o differential): Within 1 hour of arrival to lab. No STAT availability for differential.
Routine: Within 8 hours of arrival to lab
Stat Availabilty Yes
Stat Availabilty

CBC (w/o differential): Within 1 hour of arrival to lab.
No STAT availability for differential.

Performing Bench UH, WCH Hematology
Methodology/Method Description Sysmex – Hydrodynamic focusing (DC detection), flow cytometry method(semiconductor laser), SLS hemoglobin method, slides are performed manually if indicated by results.

Special Handling

Room Temp

Additional Information/Important Notes

Note: The Sysmex instrument evaluates >8000 WBC events when a CBC Auto is ordered vs CBC Man is performed on a 100 WBC count

CPT Codes

85007, 85027

Reflex Testing

Pathology QA reviews, EDTA Plt Phenom