Test Code ETOH Ethanol Level
Aliases
Alcohol Level, Ethyl Alcohol
Patient Preparation
No special patient preparation required
Special Collection Instructions
IMPORTANT NOTE: DO NOT use alcohol or other volatile disinfectants at the site of venipuncture. Aqueous Zephiran (benzalkonium chloride), aqueous Merthiolate (thimerosal), or providone-iodine may be used.
Collection Method
Venous collection
Minimum Volume
0.6 mL
Container
Lt Green / Plasma Separator / Lithium Heparin
Preferred Specimen
Light Green Top (lithium heparin) plasma
Acceptable Specimens
Serum:
- Gold Top SST
- Red Top
Plasma:
- Light Geen Top (lithium heparin)
- Purple(K2 EDTA)
Reference or Target Ranges
<10 mg/dL
Reportable Units
mg/dL
Critical Value
>300 mg/dL
Reasons for Rejection
<!-- Rejection-->
| Hemolysis Threshold | 1000 mg/dL | 
| Icterus Threshold | 60 mg/dL | 
| Lipemia Threshold | 3000 mg/dL | 
| Other | QNS, contaminated, improperly Labeled, improper specimen type | 
Specimen Stability
<!-- Stability -->
| Ambient | 2 hours once opened and recapped, 2 days capped | 
| Refrigerated (4°C to 8°C) | 2 weeks capped | 
| Frozen(-40°C to 0°C) | 4 weeks | 
Performance Information
<!-- Performance information -->
| Days and Time Performed | 24/7 | 
| Expected Turn Around Time | Stat: 1 hour Routine: 4 hours | 
| Stat Availability | Yes | 
| Performing Bench | UH Chemistry | 
| Methodology/Method Description | Siemens Atellica – Colorimetric | 
Additional Information/Important Notes
Only good 2 hours once opened and recapped.
CPT Codes
82077
 
           Home
Home Help
Help