Sign in →

Test Code Fl Amy Fluid, Amylase Level

Patient Preparation

No special patient preparation required

Special Collection Instructions

Source must be noted

Collection Method

Sterile body fluid collection

Minimum Volume

  • Adults: 0.4 ml
  • Pediatric: 0.25 ml

Container

Sterile Urine Cup

Preferred Specimen

Sterile container

Acceptable Specimens

Body fluids

Reference or Target Ranges

No reference values available for body fluid sources

Reportable Units

U/L

Critical Value

Not defined for this assay

Reasons for Rejection

 


<!-- Rejection-->

 

Hemolysis Threshold 500 mg/dL
Icterus Threshold 30 mg/dL
Lipemia Threshold 500 mg/dL
Other QNS, contaminated, improperly Labeled, improper specimen type

Specimen Stability

 


<!-- Stability -->

 

Ambient 7 days
Refrigerated (4°C to 8°C) 1 month
Frozen(-40°C to 0°C) Not listed

Performance Information

 


<!-- Performance information -->

 

Days and Time Performed 24/7
Expected Turn Around Time Stat: 1 hour
Routine: 4 hours
Stat Availabilty Yes
Performing Bench UH Chemistry, WCH Chemistry
Methodology/Method Description Siemens Atellica – Colorimetric assay

CPT Codes

82150