Sign in →

Test Code IGGS4 IgG4, Immunoglobulin Subclasses, Serum

Reporting Name

IgG4, Ig Subclasses

Useful For

Supporting the diagnosis of IgG4-related disease

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum


Ordering Guidance


This test only quantitates the IgG4 protein. If quantitation of all IgG subclass types is desired, order IGGS / IgG Subclasses, Serum.



Specimen Required


Patient Preparation: Fasting preferred but not required

Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL

Collection Instructions: Centrifuge and aliquot serum into a plastic vial.


Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Refrigerated (preferred) 14 days
  Ambient  14 days
  Frozen  14 days

Reference Values

0-<5 months: ≤19.8 mg/dL

5-<9 months: ≤20.8 mg/dL

9-<15 months: ≤22.0 mg/dL

15-<24 months: ≤23.0 mg/dL

2-<4 years: ≤49.1 mg/dL

4-<7 years: ≤81.9 mg/dL

7-<10 years: 1.0-108.7 mg/dL

10-<13 years: 1.0-121.9 mg/dL

13-<16 years: ≤121.7 mg/dL

16-<18 years: ≤111.0 mg/dL

≥18 years: 2.4-121.0 mg/dL

Day(s) Performed

Monday through Friday

Test Classification

This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.

CPT Code Information

82787

LOINC Code Information

Test ID Test Order Name Order LOINC Value
IGGS4 IgG4, Ig Subclasses 2469-5

 

Result ID Test Result Name Result LOINC Value
IGGS4 IgG4, Ig Subclasses 2469-5

Method Description

The determination of the soluble antigen concentration by turbidimetric methods involves the reaction with specific antiserum to form insoluble complexes. When light is passed through the suspension formed a portion of the light is transmitted and focused onto a photodiode by an optical lens system. The amount of transmitted light is indirectly proportional to the specific protein concentration in the test sample.

 

Concentrations are automatically calculated by reference to a calibrations curve stored within the instrument.(Package inserts: Optilite IgG4. The Binding Site Group, Ltd: ver.23, 07/2019)

Report Available

1 to 2 days

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus OK

Method Name

Turbidimetry

Secondary ID

84250