Test Code MMAS Methylmalonic Acid, Quantitative, Serum
Reporting Name
Methylmalonic Acid, QN, SUseful For
Evaluating children with signs and symptoms of methylmalonic acidemia using serum specimens
Evaluating individuals with signs and symptoms associated with a variety of causes of vitamin B12 (cobalamin) deficiency
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
SerumSpecimen Required
Collection Container/Tube:
Preferred: Serum gel
Acceptable: Red top
Submission Container/Tube: Plastic vial
Specimen Volume:1.5 mL
Collection Instructions: Centrifuge and aliquot serum into a plastic vial.
Specimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 48 days | |
Ambient | 48 days | ||
Frozen | 48 days |
Reference Values
≤0.40 nmol/mL
Day(s) Performed
Monday through Friday
Test Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
83921
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
MMAS | Methylmalonic Acid, QN, S | 13964-2 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
80289 | Methylmalonic Acid, QN, S | 13964-2 |
Method Description
Methylmalonic acid (MMA) is determined by liquid chromatography tandem mass spectrometry (LC-MS/MS) stable isotope dilution analysis. The specimen is mixed with an internal standard (methyl-d3-malonic acid). MMA and d3-MMA are isolated by solid phase extraction. LC-MS/MS is performed using mobile phases and a short column to separate MMA and d3-MMA from the bulk of the specimen matrix. The MS/MS is operated in the multiple reaction monitoring (MRM) negative mode to follow the precursor to product species transitions. Separation of MMA/d3-MMA from the more physiologically abundant succinic acid is accomplished by the careful selection of MRM transitions and optimization of the LC separation. The ratios of the extracted peak areas of MMA to d3-MMA determined by LC-MS/MS are used to calculate the concentration of MMA present in the sample.(Unpublished Mayo method)
Report Available
3 to 5 daysReject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Method Name
Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)
Forms
If not ordering electronically, complete, print, and send 1 of the following forms with the specimen:
-General Request (T239)
-Benign Hematology Test Request (T755)