Test Code TEG - Bleed TEG Bleeding
Aliases (synonyms)
TEG - Global Hemostasis with Lysis
Useful For
Evaluation of global Hemostasis:
CRT-MA represents the maximal combined platelet and fibrin clot strength.
CFF-MA measures the functional fibrinogen contribution.
Delta (CRT-MA , CFF-MA) represent the platelet contribution to clot strength.
Ly30 is increased in hyperfibrinolysis.
Patient Preparation
Avoid tissue contaminants or interfering substances, Using a 21 gauge needle or larger, venipuncuture samples - draw 2 blue tops and discard the first. Line draws discard the first 10mL of blood.
Special Collection Instructions
Unspun specimen must arrive within 1 hour of draw.
Collection Method
Use of 21 gauge or larger. Full tube. Mix well by inversion.
Minimum Volume
Tube filled to Proper 9:1 ratio
Container
Light Blue Sodium Citrate
Preferred Specimen
2.7 mL Lt. Blue (NaCitrate) vacutainer
Reference or Target Ranges
Global Hemostasis with Lysis Cartridge - Citrated K, RT, FF (07-605-US) | ||||
CK-R (min) | CK LY30 (%) | CRT-MA (mm) | CFF-MA (mm) | |
Normal Patient | 4.6-9.1 | 0.0-2.6 | 52-72 | 15-32 |
Critical Value
Not defined for this assay
Reasons for Rejection
Hemolysis Threshold | N/A |
Icterus Threshold | N/A |
Lipemia Threshold | N/A |
Patient samples must be obtained through a 21-gauge needed or larger. Underfilled , overfilled or clotted sample as well as samples that are refrigerated or centrifuged mus all be redrawn. |
Specimen Stability
Ambient | 2 hours |
Refrigerated (4°C to 8°C) | N/A |
Frozen(-70°C to -20°C) | N/A |
Performance Information
Days and Time Performed | 24/7 |
Expected Turn Around Time |
Stat: Within 60 minutes of arrival in the lab Routine: Within 2 hours on arrival to lab |
Stat Availability | Yes |
Performing Bench | UH Coagulation |
Methodology/Method Description | The TEG6s analyzer records the kinetic changes in a sample of anticoagulated whole blood as the sample clots, retracts and/or lyses in real time. The system output consists of a table of numerical values derived from a graphical representation of the hemostasis curve,which depicts clot firmness over time. |
CPT Codes
85347, 85384, 85576, 85396 26