Test Code TPO Ab Thyroid Peroxidase Antibody
Patient Preparation
No special patient preparation required
Collection Method
Venous collection
Minimum Volume
- Adults: 1 mL
- Pediatric: 0.6 ml
Container
Lt Green / Plasma Separator / Lithium Heparin
Preferred Specimen
Light Green Top (lithium heparin) plasma
Acceptable Specimens
Serum:
- Gold Top SST
- Red Top
Plasma:
- Light Geen Top (lithium heparin)
Reference or Target Ranges
<34 IU/mL
Reportable Units
IU/mL
Critical Value
Not defined for this assay
Reasons for Rejection
<!-- Rejection-->
Hemolysis Threshold | 800 mg/dL |
Icterus Threshold | 66 mg/dL |
Lipemia Threshold | 1500 mg/dL |
Other | QNS, contaminated, improperly Labeled, improper specimen type |
Specimen Stability
<!-- Stability -->
Ambient | 8 days |
Refrigerated (4°C to 8°C) | 8 days |
Frozen(-40°C to 0°C) | 24 months |
Performance Information
<!-- Performance information -->
Days and Time Performed | 24/7 |
Expected Turn Around Time | Routine: 4 hours |
Stat Availability | No |
Performing Bench | UH Chemistry |
Methodology/Method Description | Roche Cobas – Two step competition assay |
Additional Information/Important Notes
The autoantibodies are heterogeneous and this gives rise to non-linear dilution phenomena for individual samples. This assay may not be used in correlation with other manufacturers or vendors due to lack of comparibility. This assay was performed on the Roche Cobas 801 system.
CPT Codes
86376